The Model 202 maintains a constant flow rate as long as the upstream (inlet) pressure is held constant during downstream pressure fluctuations from column switching, temperature programming, sequence reversal, or backflushing — even if the fluctuation is to within 10 psi of the supply pressure.A user-variable span adjustment permits it to be used for a variety of flow ranges. The flow range is infinitely adjustable from a minimum flow as small as 5.0 mL/min up to a maximum flow of 1.6 L/min (N2 at 40 psi). After the span is adjusted, the flow controller has a full 10 turns of resolution between zero flow and the maximum rate. When equipped with a Spectrol digital dial, settings are reproducible to better than 1%.The Model 202 comes standard with 1/8" NPT female pipe threads. Adapters to go from pipe to 1/16" OD tubing are included. Other fittings are availableNOTE: With all VICI Valco flow controllers, the inlet pressure must exceed the outlet pressure by 10 psi.