Are You Properly Handling and Maintaining Your Septum? 

Restek has some suggestions....

Septum Handling

All GC septa, regardless of their composition, puncturability, or resistance to thermal degradation, will be a source of problems if they are mishandled. Always use clean forceps, or wear clean powderless or cotton gloves when handling septa. Do not handle them with bare fingers or with powdered latex gloves – contaminants such as finger oils, perfumes, make-up, fingernail polish, skin creams, hand soaps, and talcum can be absorbed into the septum and will bleed out during your analyses.

Also, always follow septum and instrument manufacturers’ recommendations when installing a septum. Overtightening a septum nut will invariably reduce septum lifetime by increasing coring and splitting problems.

Septum Bleed

All GC septa contain various amounts of volatile materials (e.g., silicone oils, phthalates) that can be released when the septum is heated to analysis temperatures. GC septum bleed occurs when these volatiles enter the column, then elute from the column, creating elevated baselines (for isothermal analyses), baseline disturbances, or extraneous (but consistent) peaks in the chromatogram. This problem is prevalent in temperature-programmed analyses because the septum volatiles collect on the column during the oven cooldown and initial hold periods.

To avoid GC septum bleed, either condition your septum prior to running your analyses, or use a preconditioned septum that is ready for immediate use. Allowing the septum to condition at operating temperatures for a few hours is an excellent way to assure optimum performance.

Septum Coring

Most modern gas chromatographs have a septum purge flow that is designed to flow across the inward face of the septum, sweeping most of the volatiles away from the column. What happens when either the septum has been punctured too many times, or the syringe is damaged or has the wrong tip type? In these cases, small septum particles can be cored from the body of the septum and fall into the inlet liner. Once in the liner, they are typically subjected to even higher temperatures than in the septum housing, and the volatile compounds they release are swept directly into the column or out the split vent.

To prevent GC septum coring, don’t overtighten the septum nut, be sure to routinely replace your septum, and inspect your syringe (manual or autosampler) for tip damage. Also, consider switching to a softer, more pliable septum. Softer septa are less likely to core than firmer septa. Remember, however, that softer septa usually have a lower maximum operating temperature than firmer septa, so consider your method requirements carefully before deciding to switch. Changing syringe needle styles can also help reduce septum coring. For example, Point Style 2 needles (beveled needle point) are much more likely to core septa (especially when the tip has become bent or dull) than a Point Style 5 needle (conical needle with side port).

Have You Chosen The Right GC Septum For the Job?

Why Are Low-Bleed Septa Important?

Either baseline rise or extraneous peaks caused by GC septum bleed can interfere with identification and quantification of target analytes.

Why Does Septum Puncturability Matter?

A septum that can be penetrated cleanly and easily by a syringe needle has a longer life. Moreover, consistent injections made through such a septum help ensure accurate results. The soft silicone rubber from which all Restek septa are manufactured is specially formulated for chromatographic performance, which ensures their septa are easy to puncture. However, in cases in which a small degree of pliability is sacrificed for high-temperature optimization, the CenterGuide dimple will help guide the syringe, for clean, consistent injections, minimizing septum coring.

What Septum Configurations Are Available, and for Which GCs?

Restek has fashioned septa for all major brands of gas chromatographs and injectors. Use Restek's handy septum size chart below to determine the septum diameter for your instrument. 

Instrument Septum Diameter
Agilent
5880A, 5890, 6850, 6890, 7890, PTV 11 mm
5700, 5880 9.5/10 mm
On-Column Injection 5 mm
Bruker/Varian (Injector Type)
Packed Column 9.5/10 mm
1078/1079 10/11 mm
1177 9 mm
1075/1077 11 mm
Finnigan (TMQ)
GC 9001 9.5 mm
GCQ 9.5 mm
QCQ 9.5 mm
TRACE 2000 9.5 mm
Gow-Mac
6890 Series 11 mm
All Other Models 9.5 mm
PerkinElmer
Sigma Series 11 mm
900, 990 11 mm
8000 Series 11 mm
Auto SYS, Auto SYS XL 11 mm
Clarus 590/690 11 mm
Pye/Unicam
All Models 7 mm
Shimadzu
All Models Plug
SRI
GCs With Heated Inlets 9.5 mm
All Other Models Plug
Thermo Scientific
TRACE 1300/1310, 1600/1610 Series 11 mm
TRACE Ultra GC 17 mm
FOCUS GC 17 mm
GCQ w/TRACE, PTV 12.7 mm
8000 Series 17 mm
Tracor
540 11.5 mm
550, 560 9.5 mm
220, 222 12.7 mm

Which Septa Should I Use?

Thermolite Plus septa are a proven low-bleed champion. With a maximum temperature of 350°C, Thermolite Plus septa are suitable for most applications.

BTO septa are bleed and temperature optimized with a maximum temperature of 400°C, for the most demanding GC and GC-MS applications. They attain remarkable softness and puncturability at high temperatures. However, at lower temperatures, BTO septa have limited puncturability and septum coring or needle damage may occur. The CenterGuide can help reduce coring when tapered (rounded tip) needles are used to puncture the septum.

Restek septa are precision molded for a consistent and accurate fit, they are preconditioned so they are ready for immediate use, and they are packaged in
ultra-clean blister packs for a reliably clean product.

Still Have Questions?

You can give us a call or email us and we can find the right GC Injection port septa for your application. It's our pleasure to help you and it's free!

1-800-267-8103 or tech@chromspec.com

Chromatographic Specialties Inc.

Welcome to the New Site!

Important Update

Welcome to the new home of MJS BioLynx! If you are a registered user, your Account Number will remain the same and you can login using your email, which will be sent a verification code to access the site. You can login using the button below. If you do not have a registered account, please click “Register” below to create your account and gain access to your pricing. If you have any questions about navigating our new site, please feel free to contact sales@chromspec.com.

User Guide

English Guide French Guide
Notice
×
Checking permissions...