ZPure™ Glass Oxygen Traps Instructions For Use
How Do I Safely and Properly Use the ZPure™ Glass O2 Filters

Pure gas is critical requirement in gas chromatography, spectroscopy, optics, lithography, and numerous other applications in manufacturing and analytical laboratories. The ZPure™ O2 filters remove O2 from He, Ar, N2, H2 and methane to low ppb levels with very high capacity.
Trajan has produced the following information and instructions.
To safely use these filters warnings must be read carefully and understood. Improper use of this product can cause harm or death to personnel and damage to property!
- Wear eye protection and use caution when working with pressurized systems.
- Maximum pressure is 11 bar / 160 psi.
- When using a high-pressure gas source over-pressure protection must be provided.
- Do not open filter, even after use.
- The rapid oxidation of the active adsorbent in this filter can cause the filter bed to become extremely hot, resulting in danger of fire or burn injuries. Limited exposure of the filter to air during installation does not result in rapid reaction or dangerous heat generation.
- Do not pass air or oxidizing gases in concentrations above 0.1% through this filter.
Compression Fittings
- Do not expose the protective outer plastic tube to solvents.
- Do not use without the protective outer plastic tube in place.
- Special precautions might be required when using hydrogen or hazardous gases.
-
Consult local regulations and your company’s safety procedures.
Installation For Compression Fittings
- Check the package contents for damage. Contact your supplier if any items are damaged or missing.
- The ZPure filters are delivered in the active state, filled with helium. It is ready for use, but it is recommended that the trap be purged with the gas to be purified.
- For the best connections, start with freshly-cut tubing in the gas line. 1/8” nuts should be turned 3/4 turn past finger tight. 1/4” nuts should be turned 1-1/4 turns past finger tight. .
- To connect in-line, first purge the gas line with supply gas. While maintaining a low purge rate of 10 to 30 cc/min, remove the end plugs from both ends of the trap, then attach the inlet end of the trap to the gas line.
- While the end plugs are removed from the trap during installation, a small amount of air will diffuse into the adsorbent. As long as the trap is attached to the gas line within a few minutes, the loss of adsorbing capacity will be negligible.
- Attach the outlet end of the trap.
- While maintaining a low purge rate of 10 to 30 cc/min, purge the entire system until a total of 1 to 2 liters of gas has been flushed through.
- Leak check all connections.
Product Numbers
ZPure Glass O2 Filter 46 cc 1/8” Brass C37236B
ZPure Glass O2 Filter 46 cc 1/4” Brass C37237B
ZPure Glass O2 Filter 46 cc 1/8” Stainless Steel C37236SS
ZPure Glass O2 Filter 46 cc 1/4” Stainless Steel C37237SS
These Oxygen Filters are also available with quick connect fittings.
Quick Connect
ZPure Glass O2 Filter, 1/8" Brass Quick Connect Fittings C37236BQC
ZPure Glass O2 Filter, 1/4" Brass Quick Connect Fittings C37237BQC
ZPure Glass O2 Filter, 1/8" Stainless Steel Quick Connect Fittings C37236SSQC
ZPure Glass O2 Filter, 1/4" Stainless Steel Quick Connect Fittings C37237SSQC
Still Have Questions?
You can give us a call or email us and we can help guide you through the process. It's our pleasure to help you and it's free!